Sustainable Construction
Lorem ipsum dolor...
Ultra compact and lightweight: Fits into a pocket
Features:
Specifications
Standard compliance | General-Purpose Sound Level Meter according to Japan Measurement Law JIS C 1509-1: 2005 Class 2, IEC 61672-1: 2002 Class 2 CE mark (EMC Directive 2004/108/EC), WEEE Directive |
Measurement functions Processing functions (normal mode) | Sound pressure level Lp, equivalent continuous sound pressure level Leq, sound exposure level LE, maximum sound pressure level Lmax, peak sound level LCpeak (only when peak range was selected) |
Measurement time settings | 1 min, 5 min, 10 min, 1 h |
Measurement level range (normal mode) | A-weighting: 30 dB to 137 dB, C-weighting: 36 dB to 137 dB |
Linear operation range (A-weighting, 1 kHz) | 30 dB to 137 dB |
Peak sound level measurement range | 65 dB to 140 dB |
Measurement frequency range | 20 Hz to 8 kHz |
Frequency weighting characteristics | A-weighting, C-weighting |
Time weighting characteristics | F (Fast), S (Slow) |
Level range | Wide range: 30 dB to 130 dB, peak range*: 65 dB to 130 dB *Peak range is used for peak sound level measurements |
RMS detection method | Digital processing |
Processing | Digital sampling cycle: 30.3 μs (Lp, Leq, LE, Lmax, Lpeak) |
Display (TN positive display, reflective) | Numeric display: 0.1 dB resolution Bar graph: Scale range 100 dB, 5 dB resolution (update frequency 0.1 s) Warning indicators: Over (Overload) appears at 137.4 dB (1 kHz) Under (Underload) appears from -0.6 dB of lower scale limit Battery capacity 3-stage indicator for remaining capacity |
Store function (processing results saved in internal memory, using manual store) | Number of data: 199 Stored data can be checked on recall screen |
DC output connector | DC output: 3 V (full scale), 25 mV/dB, output impedance: 50 Ω Load impedance: 10 kΩ or more |
AC monitor output connector | AC output: 1 Vrms +600 -400 mVrms (at 110 dB), overload: +2 dB Output impedance: 600 Ω, load impedance: 10 kΩ or more, frequency weighting: Z-weighting |
USB connector | For transfer of stored data to a computer, using optional USB adapter cable |
Power requirements (size AAA [IEC R03] battery x 2 Normal temperature) | Current consumption: approx. 90 mA (3 V operation) Battery life: approx. 7.5 hours (with alkaline batteries), (Wide range)approx. 2.5 hours (with manganese batteries) |
Temperature /humidity range for operation | -10 ℃ to +50 ℃, 10 % to 90 % RH (no condensation) |
Dimensions | Approx. 120 mm (H) × 63 (W) × 23.5 mm (D) |
Weight | Approx. 105 g (including batteries) |